Service Instruction Sheet
Introduction
These instructions are intended for use with standard comparator rebuild kit (PN 4138) or the
Skydrol rebuild kit (PN 4158). Both rebuild kits contain all of the parts and tools required to rebuild
the GaugeCalXP pressure comparator. Rebuilds should be performed only by qualied personnel.
Following service instructions are for the standard comparator rebuild kit (PN 4138). If rebuilding the Skydrol
GaugeCalXP, cross reference the appropriate o-ring part numbers in the Parts List located on the last page.
Service Instructions
1 Remove all gauges and ttings from the
GaugeCalXP pressure comparator (gure 1).
2 Tip the comparator on its end to completely
drain the water or oil from the system.
Always dispose of uids in a responsible
manner.
3 Use a 3mm Allen key to remove the two (2)
cap screws securing the Reservoir to the
Main Seal Block, then lift to remove the Res-
ervoir and Reservoir o-ring (PN 2018)
(gure 2).
4 Remove the press-t Reservoir Vessel,
and replace the lower seal o-ring (PN 3303
or PN 3376) (gure 2).
*
Note:The GaugeCalXP is assembled with
a Reservoir o-ring in one of two sizes. Com-
pare your o-ring to the illustrated parts list
on page 6, to determine the appropriate
replacement size for your unit.
5 Use a 4- or 5mm Allen key to remove the
Handle Assembly bolt and washer(s), then
pull to slide the Handle Assembly from the
Piston Assembly (gure 3). Be sure to also
remove the Key from the Piston Assembly
Leadscrew (gure 3).
PRESSURE COMPARATOR
10000PSI–700bar
Gau
g
eCa
lxp
Figure 1
Figure 2
Figure 3
C R Y S T A L
engineering corporation
PSI
50
250
0 300
100
Gauge
Base
Reservoir
O-ring
(PN 2018)
O-ring
(PN 3303)
(PN 3376)
or
*
Handle
Key
Leadscrew
DESIGNED FOR USE WITH CRYSTAL PRESSURE FITTINGS
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - Service Instruction Sheet

Service Instruction SheetIntroductionThese instructions are intended for use with standard comparator rebuild kit (PN 4138) or the Skydrol rebuild ki

Page 2 - C R Y S T A L

Page 2 • GaugeCalXP Service Instruction SheetC R Y S T A L engineering corporation6 Turn the comparator over to gain access to the underside of t

Page 3

GaugeCalXP Service Instruction Sheet • Page 3C R Y S T A L engineering corporation12 To remove the three (3) stainless steel Guide Rods, insert t

Page 4

Page 4 • GaugeCalXP Service Instruction SheetC R Y S T A L engineering corporation16 Remove the MPF-1/4FPT and MPM-MPM t-tings from the Rear Gau

Page 5

GaugeCalXP Service Instruction Sheet • Page 5C R Y S T A L engineering corporation23 From the front of the Fine Adjust body, remove the o-ring (P

Page 6 - Sizes to Scale

Page 6 • GaugeCalXP Service Instruction SheetParts ListO-ring 3303 / 416634.65mm x 1.78mm1.364in x 0.070inO-ring 3376 / 416735mm x 2.5mm1.378in x

Page 7

GaugeCalXP Service Instruction Sheet • Page 7C R Y S T A L engineering corporationHow to Contact Us:Phone (805) 595-5477Toll-Free (800) 444-1850

Page 8

C R Y S T A L engineering corporation© 2010 Crystal Engineering Corporation708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701PN 4139 Rev