Operation Manualfor XP2i Digital Test Gauge (Standard and Dual-Display(DD))
Functions 8XP2i Operation ManualResetIf for some reason the XP2i needs to be reset, remove any battery for at least one minute, then reinstall the
Functions 9XP2i Operation ManualInternal Data Logging (-DL DataLoggerXP Option)With the DataLoggerXP firmware update option you can record up to six
Enclosure 10XP2i Operation ManualEnclosureAluminum alloy, designed to meet IP67 ratings.Weight: 562 g (19.8 oz), including batteries. ALL DIMENSIONS
Functions 11XP2i Operation ManualXP2I SERIAL NUMBERSSerial Number LocationThe serial number of your XP2i is located on the stem above the pressur
Specifications 12XP2i Operation ManualSpecificationsAccuracyIncludes all effects of linearity, hysteresis, repeatability, temperature, and stability fo
Specifications 13XP2i Operation ManualMedia CompatibilityLiquids and gases compatible with 316 Stainless Steel.Pressure Conversions1 PSI = 27.6806 i
Specifications 14XP2i Operation ManualPOWERBatteries . . . . . . . . . . . . . . . . . . . . . . . . Three size AA (LR6) batteries. !WARNING: Do no
Specifications 15XP2i Operation ManualPRESSURE RANGES, DISPLAY SCALES, & RESOLUTIONPSI bar kPa Overpressure PSIinH20 inHg mmHg mmH20 kg/cm2bar
Safety & Certifications 16XP2i Operation ManualSafety & CertificationsHAZARDOUS LOCATIONSEvery XP2i pressure gauge includes the following In
Safety & Certifications 17XP2i Operation ManualEMC EC Declarations of Conformity ATEX EC Declarations of ConformityEC Declaration of Conformity I
ContentsOverview ... 1Introduction ...
Safety & Certifications 18XP2i Operation ManualPED EC Declarations of ConformityI/We of declare that EC Declaration of Conformity Crystal Enginee
Safety & Certifications 19XP2i Operation ManualATEX/IECEX SAFETY INSTRUCTIONSBezpečnostní Instrukce Pro Prostředí s Nebezpečím Výbuchu – ČESKY
Safety & Certifications 20XP2i Operation ManualSafety Instructions for Hazardous Locations – ENGLISH (English)• Do not use the RS-232 connector
Safety & Certifications 21XP2i Operation ManualInstructions De Sécurité Pour Les Zones Dangereuses – FRANÇAIS (French)• Ne pas utilisez le conne
Safety & Certifications 22XP2i Operation ManualVeiligheidsinstructie Voor Gebruik In Een Explosie Gevaarlijke Omgeving – NEDERLANDS (Dutch)• Het
Safety & Certifications 23XP2i Operation ManualRäjähdysvaarallisten Tilojen Turvallisuusohjeita – SUOMEN KIELI (Finnish)• RS-232 väylää/liitintä
Support 24XP2i Operation ManualSupportTROUBLESHOOTINGThe XP2i is a very high performance gauge. Due to the high resolution of this product, you m
Support 25XP2i Operation ManualCALIBRATIONIf adjustment is required, we recommend returning the unit to the factory. Factory service offers benefit
Support 26XP2i Operation ManualSOFTWARELabVIEW™ Drivers and Programming InstructionsControl and communicate with XP2is using National Instrument’
Support 27XP2i Operation ManualCONTACT USPhone . . . . . . . . . . . . . . . . . . . . . . . . . .(805) 595-5477Toll-Free . . . . . . . . . . .
Overview 1XP2i Operation ManualOverviewINTRODUCTIONThank you for choosing the XP2i Digital Test Gauge from Crystal Engineering Corporation. Your
4661.C1© 2013 Crystal Engineering Corporation708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701
Overview 2XP2i Operation ManualOPERATING INSTRUCTIONSThe XP2i is shipped with batteries installed, so it’s ready to use. Press and hold the (on/o
Functions 3XP2i Operation ManualFunctions Units ButtonPressing this button causes the XP2i to select the next available unit of pressure measureme
Functions 4XP2i Operation ManualTare (-DD, Dual-Line Display XP2i Only)Tare is a constant value subtracted from the true pressure. For instance, if
Functions 5XP2i Operation Manual Peak Detection and AverageOn the XP2i, pressing the Peak button causes the display to cycle through the following:
Functions 6XP2i Operation ManualAutomatic Shut-oThe XP2i has a shutoff timer and will turn off automatically after 20 minutes of non-operation. Pres
Functions 7XP2i Operation ManualWater Density (Inches of Water)The following applies only to models where inch of water is a selectable pressure un