Operation 2
GaugeCalXP Operation Manual
Operation
Follow these instructions for safe and reliable operation of your GaugeCalXP Pressure Comparator.
1 Mount the c
omparator to your bench or table: At higher pressures the force required to rotate the handle may cause the base of the comparator to lift.
Baseplate
16.25
(0.34)
19.4
(0.76)
Ø 10.00 (Ø 0.39), 4 places
Maximum bolt size: 8mm or 5/16"
174.9
(6.88)
142
(5.60)
390
(15.35)
428.8
(16.88)
All dimensions in mm (in)
Baseplate dimensions and suggested bolt sizes
2 Install the CPF-compatible  ttings and any applicable adapters:
Crystal Pressure Fittings (CPF) simplify your connections and create fi nger-tight, leak-free connections up to 10 000 PSI/700 bar/70MPa working pressure.
No sealing tape is required. See the diagram below, or visit the website, for more information on CPF.
Y
our GaugeCalXP is fi tted with a CPF base (7/16-20 MP) and will connect with the two included CPF male fi ttings (MPM-MPM) and two included 1/4" female
ttings (MPF-1/4FPT). Should you have a need for additional fi ttings, visit crystalengineering.net or contact your Crystal Distributor. Refer to the diagram
on the f
ollowing page for connection information.
Finger-tight seal uses a
Viton 90 o-ring for a seal up to
10 000 PSI / 700 bar/ 70MPa
(-20 to 50
o
C)
Wrench-tight creates a
metal/metal seal up to
10 000 PSI / 700 bar/ 70MPa
(-40 to 150
o
C)
point of contact