Con g Function Tab 8
CrystalControl Operation Manual
Modules: Temperature Module (RTD100)
The following features are unique to the Temperature Modules (RTD100):
•
Temperature Coe cient of Resistance (TCR): Select your appropriate platinum RTD Temperature Coe cient of Resistance (TCR) from the pull down menu.
When selected, the pre-populated Callendar-Van Dusen Coe cients will appear. See the nVision Operation Manual for further details on RTDs and the
RTD100 module.
•
Base Resistance (Ro): Adjust your RTD resistance at 0°C. Factory default is 100Ω.
•
Lead Type: Adjust your RTD sensor wiring between 2-, 3-, or 4-wire methods.
Typical view of RTD100 Module features in CrystalControl.